Possible
Causes and Remedies
Cannot
Set Zero Unit |
|
|
Not using
Reagent Dye Solution for protein mode. |
|
Air
bubble(s) in cuvet. |
|
Lamp output
reduced or off -- check to see if lamp is on.(It
will be bright blue). If lamp is off, check
power. If the power is on and lamp is off, send
unit to UDY Corporation. |
|
Cuvet is
dirty or coated with spilled dye. |
|
Electronic
problem - send to UDY Corporation. |
Unstable
Readings |
|
|
Air
bubble(s) in cuvet |
|
Electronic
problem - send to UDY Corporation |
|
Power
supply extremely unstable or "noisy"
Use Surge Protector. |
|
Lamp about
to fail.(It will be dim blue). |
|
Reaction
time incomplete, and dye concentrations changing |
|
Solutions
have spilled onto the outside of the Cuvet. |
|
Sample-Reagent
Dye Solution not filtered adequately |
Zero
Unstable |
|
|
Zero
potentionmenter bad - call or send unit to UDY
Corporation |
|
Insufficient
warm-up time. |
|
Electronic
problem - send to UDY Corporation |
Difficulty
Setting GAIN |
|
|
Potentionmenter
worn out. Call or send unit to UDY Corporation
|
|
Wrong
solutions used to Setup Unit. |
No
Decimal Display |
|
|
Signal to
display out of display range. Check for proper
Solutions when setting up. |
|
Electronic
problem - send to UDY Corporation |
Lamp
Not On |
Replace
Lamp - Send back to UDY Corporation for factory
replacement and alignment. |
Display
On or Blank with Decimal Point On. |
Electronic
problem - send to UDY Corporation |
NO
DISPLAY |
Bad Power
Supply. Replace the Power Supply |
|
|
|